R Series

  1. Go to Supervisor Mode > WiFi.
  2. Press Read from Card.
  3. Configure as follows:
    1. Use Hostname: Enter Yes.
    2. Server Servername (Full Disclosure): Enter dxsvc.zollonline.com.
    3. Server Servername (Full Disclosure): Enter your Server DNS name. If you do not know the name, contact your IT department.
    4. Server Port (Full Disclosure): Enter 443.
    5. Server Port (Full Disclosure): Enter 444 or your DXS server port.
    6. Username (full Disclosure): Enter the DAK Alternate ID.
    7. Password (Full Disclosure): Enter the DAK password.
    8. Edit Network Profile:
      • SSID: Enter your SSID.
      • Security Key: Enter your security key.
    9. Continue pushing the ‘Next item’ button until ‘Save to WiFi Card’ displays.
    10. Select Save to WiFi Card.
    11. After save completes, push Category List.
    12. Push Main Menu.
    13. Push Exit Config.
    14. The highlighted option is set to Save Changes and Exit – push Enter.
    15. After saving is complete, set the option for Save Changes to Card to No and push Enter.
    16. Power off the R Series.
    17. Wait several seconds and turn the R Series back on.
    18. If not already configured, set Defib to Manual mode and Confirm.
    19. Select Report Data, then ‘Transfer Mode', then ‘Report to Wi-Fi’.
    20. Push Send Report and wait until you see 100%. The cases are now uploaded to the server.
X Series or AED 3 devices? If you are done, find out how to use this feature.